FDA Recall Terminated

Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.

Recall: Z-0295-2016 · Initiated October 23, 2015

Recall

Recall Number
Z-0295-2016
Event Number
72484
Firm
Fresenius Medical Care Renal Therapies Group, LLC
FEI Number
3001451489
Product Code
KPO
Status
Terminated
Root Cause
Device Design
Initiated
October 23, 2015
Terminated
March 2, 2021
Address
920 Winter St, Waltham, MA, 02451-1521

Description

Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.

Reason

Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.

Action

Fresenius issued an Urgent - Medical Device Correction notification, dated October 21, 2015, on 10/27/15. The letter identified the affected device and the reason for the recall. It also contained instructions (on how to perform the required wiring modification) along with a Reply Form. Customers are to return the form after completing the modification. Inquiries regarding the correction should be directed to Fresenius Technical Services at 800-227-2572, referencing the Field Action number FA-2015-01-W.

Distribution

Nationwide Distribution.

Quantity

348 units