FDA Recall Open, Classified

Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.

Recall: Z-0294-2022 · Initiated November 4, 2021

Recall

Recall Number
Z-0294-2022
Event Number
88953
Firm
Medtronic Neuromodulation
FEI Number
2182207
Product Code
MHY
Status
Open, Classified
Root Cause
Other
Initiated
November 4, 2021
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568

Description

Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.

Reason

The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.

Action

The recalling firm issued undated letters on 11/4/2021 via USPS Priority Shipping. The recalling firm has three different letters addressed to the Healthcare Professional depending upon the type of device the wireless recharger is used with (InterStim Micro, product number RS5200; Activa RC, product number RS6200; or RestoreUltra & RestoreSensor, Product number RS7200, which is currently not distributed in the U.S. at this time). The letter gives an issue overview and provides requested actions. Those actions include reinforcing the instructions in the enclosed Patient Quick User Guide with their patients and to attempt to reset the WR if the device appears unresponsive. If the problem is not resolved, then contact information is provided for Medtronic Technical Services. A customer confirmation form dated November 2021 was enclosed to acknowledge the consignee has received the letter.

Distribution

US Nationwide, including PR. There was government distribution but no military distribution was identified.

Quantity

6,406 devices