Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N G047-80101. For Use in the Preparation of acid concentrate for hemodialysis.
Recall
- Recall Number
- Z-0294-2016
- Event Number
- 72484
- Firm
- Fresenius Medical Care Renal Therapies Group, LLC
- FEI Number
- 3001451489
- Product Code
- KPO
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 23, 2015
- Terminated
- March 2, 2021
- Address
- 920 Winter St, Waltham, MA, 02451-1521
Description
Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N G047-80101. For Use in the Preparation of acid concentrate for hemodialysis.
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Fresenius issued an Urgent - Medical Device Correction notification, dated October 21, 2015, on 10/27/15. The letter identified the affected device and the reason for the recall. It also contained instructions (on how to perform the required wiring modification) along with a Reply Form. Customers are to return the form after completing the modification. Inquiries regarding the correction should be directed to Fresenius Technical Services at 800-227-2572, referencing the Field Action number FA-2015-01-W.
Nationwide Distribution.
918 units