FDA Recall Terminated

Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0)

Recall: Z-0294-2008 · Initiated October 17, 2007

Recall

Recall Number
Z-0294-2008
Event Number
45492
Firm
Medtronic Navigation, Inc
FEI Number
1000517638
Product Code
JZL
Status
Terminated
Root Cause
Software change control
Initiated
October 17, 2007
Posted
November 27, 2007
Terminated
November 27, 2007
Address
826 Coal Creek Circle, Louisville, CO, 80027-9710

Description

Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0)

Reason

Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)

Action

Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.

Distribution

Worldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany

Quantity

7 units