FDA Recall
Terminated
Terumo Advanced Perfusion System 1 Electronic Oxygen Blender/analyzer; Catalog number 801188.
Recall: Z-0294-2007
·
Initiated January 1, 2004
Recall
- Recall Number
- Z-0294-2007
- Event Number
- 36393
- Firm
- Terumo Cardiovascular Systems Corp
- FEI Number
- 1828100
- Product Code
- DTQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 1, 2004
- Posted
- December 19, 2006
- Terminated
- February 2, 2007
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103-9586
Description
Terumo Advanced Perfusion System 1 Electronic Oxygen Blender/analyzer; Catalog number 801188.
Reason
The flow meter could malfunction resulting in loss of touchscreen control of the gas system, but the alternate mechanical control will remain operable.
Action
U.S. consignees were visited by firm representatives between January 2004 and September 2005 and the device modified to correct the problem. Approximately 32 exported units are still in need of being upgraded.
Distribution
Worldwide, including USA, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Quantity
199 units