FDA Recall Terminated

55CM DECATHLON STD KIT Models: DE50SH55 and DE50SH55-K 15.5 Fr Polyurethane twin lumen hemodialysis catheter with separated tips, cuff and side holes for chronic hemodialysis and apheresis.

Recall: Z-0293-2010 · Initiated October 23, 2009

Recall

Recall Number
Z-0293-2010
Event Number
53605
Firm
Spire Biomedical, Inc. One Patriots Park
FEI Number
1000159376
Product Code
MSD
Status
Terminated
Root Cause
Process control
Initiated
October 23, 2009
Posted
November 17, 2009
Terminated
June 9, 2010
Address
Bedford, MA, 01730-2343

Description

55CM DECATHLON STD KIT Models: DE50SH55 and DE50SH55-K 15.5 Fr Polyurethane twin lumen hemodialysis catheter with separated tips, cuff and side holes for chronic hemodialysis and apheresis.

Reason

There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira

Action

Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.

Distribution

Worldwide distribution - Nationwide, Government. Foreign: Canada, France.

Quantity

828 units