FDA Recall Terminated

IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and list 2220-20 (100 tests - reagents alone); Abbott Laboratories, Abbott Park, IL 60064 USA

Recall: Z-0293-04 · Initiated November 23, 2003

Recall

Recall Number
Z-0293-04
Event Number
27835
Firm
Abbott Laboratories HPD/ADD/GPRD
FEI Number
1415939
Product Code
CGN--
Status
Terminated
Root Cause
Other
Initiated
November 23, 2003
Posted
January 6, 2004
Terminated
July 22, 2004
Address
100/200 Abbott Park Road, Abbott Park, IL, 60064

Description

IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and list 2220-20 (100 tests - reagents alone); Abbott Laboratories, Abbott Park, IL 60064 USA

Reason

Chemotherapeutic concentrations of methotrexate in patient samples may carryover to subsequent samples, which would result in falsely elevated folate values, when performing IMx Folate testing.

Action

Abbott notified their international affiliates via e-mail on 11/23/03. Important Product Information letters dated 11/24/03 will be sent to the end users by the affiliates, informing them of the possible methotrexate carryover to subsequent samples and falsely elevated folate values.

Distribution

There was no U.S. distribution. The product was internationally distributed through Abbott subsidiaries in Colombia, Chile, Uruguay, Argentina, Brazil, Canada, England and Australia.

Quantity

3,475 packs