FDA Recall Open, Classified

IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment.

Recall: Z-0292-2018 · Initiated August 23, 2017

Recall

Recall Number
Z-0292-2018
Event Number
78736
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
MHX
Status
Open, Classified
Root Cause
Software design
Initiated
August 23, 2017
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment.

Reason

Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors

Action

Philips sent an Urgent Medical Device Correction letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips will provide software update and is provding interim instructions for the operator. For questions contact your local Philips representative.

Distribution

Worldwide Distribution and US Nationwide

Quantity

7071