FDA Recall Terminated

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351 Worldwide Except USA Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 Exchange part: 453564262531 453564262551 453564262571 453564262591 453564615351 453564615371 453564467761 (Korea) 453564467781 (Korea) 453564451791 (Japan, Taiwan, Indonesia) 453564451811 (Japan, Taiwan, Indonesia) 453564467821 (Mexico) 453564467841 (China) 453564467861 (China) Worldwide except for the USA SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

Recall: Z-0292-2017 · Initiated September 19, 2016

Recall

Recall Number
Z-0292-2017
Event Number
75209
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
September 19, 2016
Terminated
May 25, 2018
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351 Worldwide Except USA Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 Exchange part: 453564262531 453564262551 453564262571 453564262591 453564615351 453564615371 453564467761 (Korea) 453564467781 (Korea) 453564451791 (Japan, Taiwan, Indonesia) 453564451811 (Japan, Taiwan, Indonesia) 453564467821 (Mexico) 453564467841 (China) 453564467861 (China) Worldwide except for the USA SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

Reason

Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX

Action

The firm, Philips Healthcare, sent a "Field Safety Notice (FSN)" dated 9/14/2016 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to make sure that the MX40 Patient Wearable Monitor local configuration setting 'Alarms On' is configured to its factory default until your software is upgraded; and follow the instructions listed in the letter. A Philips Healthcare representative will contact customers with affected devices to arrange for the installation of the software upgrade. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377.

Distribution

Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.

Quantity

1824 units