FDA Recall Terminated

Battery Caps used with Animas 2020 Insulin Pump. The pump is used to help maintain blood glucose targets.

Recall: Z-0292-2009 · Initiated August 4, 2008

Recall

Recall Number
Z-0292-2009
Event Number
49476
Firm
Animas Corporation
FEI Number
3002920072
Product Code
LZG
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 4, 2008
Posted
November 25, 2008
Terminated
February 4, 2010
Address
200 Lawrence Dr, West Chester, PA, 19380

Description

Battery Caps used with Animas 2020 Insulin Pump. The pump is used to help maintain blood glucose targets.

Reason

Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device to administer insulin therapy which may result in hyperglycemia.

Action

The recalling firm issued a notification letter to the patients with insulin pumps on 8/5/08 to inform them of the problem and that they needed to replace the battery. The recalling firm contacted OneTouch Ping users by telephone on 8/5/08 for this issue and another issue. The recalling firm issued letters to distributors informing them of the problem and notified customers to replace the batteries. In addition, a letter was issued to physicians to alert them of the situation with regard to current patients. Consumers can contact 1-800-423-4087 for additional information or questions.

Distribution

Worldwide Distribution including USA (AR, CA, FL, KY, LA, MA, NE, OH, PA, SC, SD, TN, and UT) and countries of Australia, Austria, Canada, Czech Republic, Finland, France, Israel, Italy, New Zealand, Spain, Switzerland, and Sweden.

Quantity

2,899 units