Pulsox-3iA. LCD, Sp02, PR, Pulse strength bar, motion artifact, low battery message, error message, low Sp02 message, event mark. Measuring range Sp02 0 to 100, pulse rate 20 to 250, measuring accuracy Sp02 +/-2% (70 to 100), pulse rate +/- 2, Sp02 alarm, audible and visible, probe type finger, multi-site, paitent range adult to neonatal, memory function approx. 24 hrs, printer/recorder optional accessory, output RS232 w/optional interface module, power supply 2 AAA batteries, Battery life approx. 48 hrs, size 42x68x20mm, weight 42g.

Recall: Z-0290-04 · Initiated November 14, 2003

Recall

Recall Number
Z-0290-04
Event Number
27692
Firm
Minolta Corp
FEI Number
1000136179
Product Code
DQA
Status
Terminated
Root Cause
Other
Initiated
November 14, 2003
Posted
July 20, 2004
Terminated
April 14, 2004
Address
101 Williams Dr, Ramsey, NJ, 07446-1217

Description

Pulsox-3iA. LCD, Sp02, PR, Pulse strength bar, motion artifact, low battery message, error message, low Sp02 message, event mark. Measuring range Sp02 0 to 100, pulse rate 20 to 250, measuring accuracy Sp02 +/-2% (70 to 100), pulse rate +/- 2, Sp02 alarm, audible and visible, probe type finger, multi-site, paitent range adult to neonatal, memory function approx. 24 hrs, printer/recorder optional accessory, output RS232 w/optional interface module, power supply 2 AAA batteries, Battery life approx. 48 hrs, size 42x68x20mm, weight 42g.

Reason

Error in the internal software. When a pulse level is high or subject has SPO2 of less than 80%, the data shifts.

Action

Distributor provided customer list to Konica Minolta who initiated phone calls to the customers on 11/14/2003.

Distribution

The units were shipped to one distributor in Utah.

Quantity

9