Pulsox-3Si. LCD, SpO2, PR, Pulse strength bar, Motion artifact, Low battery message, Error message, Event mark, Measuring range Sp02 (%) 0 to 100, Pulse rate 20 to 250, Measuring accuracy +/- 2% (70 to 100), Pulse rate +/- 2%, Sp02 alarm, no, Probe type, finger, multi-site, Patient range adult to neonatal, memory function Approx. 24 hrs, printer/recorder optional accessory, output RS232 w/optional interface module, power supply 2 AAA batteries, Battery life approx. 48 hours, size: 42x68x20mm, weight 42g.
Recall
- Recall Number
- Z-0289-04
- Event Number
- 27692
- Firm
- Minolta Corp
- FEI Number
- 1000136179
- Product Code
- DQA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 14, 2003
- Posted
- July 20, 2004
- Terminated
- April 14, 2004
- Address
- 101 Williams Dr, Ramsey, NJ, 07446-1217
Description
Pulsox-3Si. LCD, SpO2, PR, Pulse strength bar, Motion artifact, Low battery message, Error message, Event mark, Measuring range Sp02 (%) 0 to 100, Pulse rate 20 to 250, Measuring accuracy +/- 2% (70 to 100), Pulse rate +/- 2%, Sp02 alarm, no, Probe type, finger, multi-site, Patient range adult to neonatal, memory function Approx. 24 hrs, printer/recorder optional accessory, output RS232 w/optional interface module, power supply 2 AAA batteries, Battery life approx. 48 hours, size: 42x68x20mm, weight 42g.
Error in the internal software. When a pulse level is high or subject has SPO2 of less than 80%, the data shifts.
Distributor provided customer list to Konica Minolta who initiated phone calls to the customers on 11/14/2003.
The units were shipped to one distributor in Utah.
291