FDA Recall Terminated

Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.

Recall: Z-0288-04 · Initiated October 17, 2003

Recall

Recall Number
Z-0288-04
Event Number
27490
Firm
Ohmeda Medical, A Div of Datex-Ohmeda Inc
FEI Number
1121732
Product Code
FMZ
Status
Terminated
Root Cause
Other
Initiated
October 17, 2003
Posted
July 20, 2004
Terminated
October 11, 2005
Address
8880 Gorman Rd, Laurel, MD, 20723-5800

Description

Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.

Reason

Heater vent door on neonatal incubator may malfunction.

Action

The recalling firm notified domestic consignees by certified letter and international distributors by e-mail dated 10/17/2003. The recall notice was incorporated into a firm''s follow-up letter concerning an on-going product correction of the elevating base on the same neonatal beds. The letter advises of the malfunctioning heater door vent and states that the firm will replace all existing vent doors with doors made of higher temperature rated material. Firm will provide retrofit kits for consignee or service representative installation.

Distribution

Products were sold to 365 direct accounts around the US and 58 international distributors.

Quantity

1404 beds