Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.
Recall
- Recall Number
- Z-0288-04
- Event Number
- 27490
- Firm
- Ohmeda Medical, A Div of Datex-Ohmeda Inc
- FEI Number
- 1121732
- Product Code
- FMZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 17, 2003
- Posted
- July 20, 2004
- Terminated
- October 11, 2005
- Address
- 8880 Gorman Rd, Laurel, MD, 20723-5800
Description
Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.
Heater vent door on neonatal incubator may malfunction.
The recalling firm notified domestic consignees by certified letter and international distributors by e-mail dated 10/17/2003. The recall notice was incorporated into a firm''s follow-up letter concerning an on-going product correction of the elevating base on the same neonatal beds. The letter advises of the malfunctioning heater door vent and states that the firm will replace all existing vent doors with doors made of higher temperature rated material. Firm will provide retrofit kits for consignee or service representative installation.
Products were sold to 365 direct accounts around the US and 58 international distributors.
1404 beds