FDA Recall Terminated

Hi-Art(R) System, Model Number H-000-0003 Product Usage: The TomoTherapy treatment system is intended to be used as an integrated system for t he planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissues. The megavoltage x-ray radiation is delivered in a rotational, non-rotational, modulated (IMRT), or non- modulated (non-IMRT/3 dimensional conformal) format in accordance with the physicians prescribed and approved plan.

Recall: Z-0286-2018 · Initiated November 17, 2017

Recall

Recall Number
Z-0286-2018
Event Number
78686
Firm
Accuray Incorporated
FEI Number
3003873069
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
November 17, 2017
Terminated
July 25, 2021
Address
1240 Deming Way, Madison, WI, 53717-1954

Description

Hi-Art(R) System, Model Number H-000-0003 Product Usage: The TomoTherapy treatment system is intended to be used as an integrated system for t he planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissues. The megavoltage x-ray radiation is delivered in a rotational, non-rotational, modulated (IMRT), or non- modulated (non-IMRT/3 dimensional conformal) format in accordance with the physicians prescribed and approved plan.

Reason

Uncontrolled couch Z-axis movement (descent)

Action

Accuray sent an Urgent Field Safety Notice letter dated November 8, 2017, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated the following: "Accuray(R) will correct all affected High-Performance Patient Couches. An Accuray representative will contact you to arrange a time to update your system. The update includes the installation of TomoTherapy(R)' H(TM) Series software version 2.1.2 (Hi Art(R) vS.1.2) on your system, as well as the replacement of cables within the couch subsystem." For questions contact Accuray Customer Support by phoning the regional support lines.

Distribution

Worldwide Distribution

Quantity

250 units