FDA Recall Terminated

Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Recall: Z-0283-2018 · Initiated November 15, 2017

Recall

Recall Number
Z-0283-2018
Event Number
78583
Firm
Implant Direct Sybron Manufacturing, LLC
FEI Number
3001617766
Product Code
NDP
Status
Terminated
Root Cause
Labeling mix-ups
Initiated
November 15, 2017
Terminated
October 9, 2018
Address
3050 E Hillcrest Dr, Westlake Village, CA, 91362-3171

Description

Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Reason

The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.

Action

On November 22, 2017, the recalling firm will send consignees customer letters with replacement product, requesting nonconforming product to be returned to Implant Direct with return labels and packaging. The recalling firm will then initiate calls 24 hours after the communication is sent to the clinician; initiate second calls 72 hours after the communication is sent; and initiate third calls 5 working days after the communication is sent. For further questions, please call (818) 444-3300.

Distribution

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.

Quantity

560 units (worldwide)