One Touch Ping Glucose Monitoring System Combines the functionality of OneTouch(R) Ping (TM) Insulin Pump and a OneTouch(R) Ping (TM) Meter Remote through radio frequency (RF) communication. Together, they provide an option to help make insulin delivery more discreet and flexible. Also, allows the most recent blood glucose results from the OneTouch(R) Ping (TM) Meter Remote to be automatically entered into bolus insulin calculations to cover carbohydrates in food or to correct for a high blood glucose level.
Recall
- Recall Number
- Z-0283-2009
- Event Number
- 49475
- Firm
- Animas Corporation
- FEI Number
- 3002920072
- Product Code
- LZG
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 4, 2008
- Posted
- November 7, 2008
- Terminated
- November 12, 2008
- Address
- 200 Lawrence Dr, West Chester, PA, 19380
Description
One Touch Ping Glucose Monitoring System Combines the functionality of OneTouch(R) Ping (TM) Insulin Pump and a OneTouch(R) Ping (TM) Meter Remote through radio frequency (RF) communication. Together, they provide an option to help make insulin delivery more discreet and flexible. Also, allows the most recent blood glucose results from the OneTouch(R) Ping (TM) Meter Remote to be automatically entered into bolus insulin calculations to cover carbohydrates in food or to correct for a high blood glucose level.
display old, inaccurate values for bolus amount delivered and amount of planned bolus totals
On 8/6/08, the recalling firm telephoned all users to inform them of problem and the need to return the system.
Nationwide Distribution
251 units