Si emans Simview NT Cassette Holder Electronic Imaging Device (Image Identifier) with CT, Part No: 54 96 984; Siemens Medical Solutions USA, Concord, CA 94520
Recall
- Recall Number
- Z-0283-2008
- Event Number
- 45495
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- KPQ
- Status
- Terminated
- Root Cause
- Labeling design
- Initiated
- February 27, 2007
- Posted
- January 31, 2008
- Terminated
- February 3, 2011
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
Si emans Simview NT Cassette Holder Electronic Imaging Device (Image Identifier) with CT, Part No: 54 96 984; Siemens Medical Solutions USA, Concord, CA 94520
Loose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
Customers were sent a Customer Safety Advisory Notice on February 27, 2007, notifying them that the mounting screws on the cassette holders on the SIMVIEW 3000 and NT Simulators, as well as some MEVASIM Simulators, could loosen and can cause damage to the system, or personal injury. The firm requested that they check of tightness on the mounting hardware. A STOP Modification notice was issued on 06/01/2007, to all affected regional service organizations that are responsible for implementing updates due to new information. Updated letters and User instructions will be released in August 2007.
Worldwide Distribution - USA including states of IN, VA, MA, VT, NJ, NY, PA, DC, NC, SC, GA, AL, TN, MI, KY, OH, IN MI, MN, ND, MT, IL, MO, KS, NE, AR, OK, TX, AZ, CA, SD, LA, Puerto Rico, and countries of Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Colombia, Costa Rica, Croatia, Denmark, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Hungary, India, Iran, Ireland, Italy, Kuwait, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, China, Israel/Palestine, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Singapore, South Africa, Spain, Thailand, Turkey, Ukraine, UK, and Vietnam.
287 units total, all recalls Z-0281-285-2008