FDA Recall Open, Classified

SED URS LP, Mobile X-ray system

Recall: Z-0282-2026 · Initiated June 10, 2025

Recall

Recall Number
Z-0282-2026
Event Number
97075
Firm
SEDECAL SA Pelaya, 9 Pol. Pol. Ind. Rio De Janeiro Algete Spain
FEI Number
3002495042
Product Code
IZL
Status
Open, Classified
Root Cause
Device Design
Initiated
June 10, 2025
Posted
October 23, 2025

Description

SED URS LP, Mobile X-ray system

Reason

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

Action

Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.

Distribution

US: CA, IL, and NJ