FDA Recall Terminated

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Recall: Z-0281-2019 · Initiated September 17, 2018

Recall

Recall Number
Z-0281-2019
Event Number
81036
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
KPA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 17, 2018
Terminated
August 21, 2020
Address
1700 Leider Ln, Buffalo Grove, IL, 60089-6622

Description

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Reason

The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.

Action

The recalling firm issued letters dated 9/3/2018 via email on 9/17/2018. A second letter dated 9/24/2018 was issued to update the acknowledgment form to include the number of affected power supply units the consignee destroyed.

Distribution

Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.

Quantity

876 devices (in U.S. and Canada)