O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GER SYS PRODUCT GER, Model Number Bl70000027GER Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Recall
- Recall Number
- Z-0279-2019
- Event Number
- 81153
- Firm
- Medtronic Navigation, Inc.-Littleton
- FEI Number
- 3004785967
- Product Code
- OXO
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 30, 2018
- Terminated
- June 18, 2020
- Address
- 300 Foster St, Littleton, MA, 01460-2017
Description
O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GER SYS PRODUCT GER, Model Number Bl70000027GER Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
Urgent Medical Device Correction notification letters were sent to customers on 8/30/18. The letter identified the affected letter, problem and actions to be taken. All consignees will received the notification for 3.1.7 software maintenance release. Field service personnel will install the software onto the designed consignee systems.
Worldwide - US Distribution in the states: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and Puerto Rico. The products were distributed to the following foreign countries: Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK, Venezuela.
596 total