FDA Recall Open, Classified

Prismaflex Control Unit, software versions below 7.21

Recall: Z-0278-2020 · Initiated October 4, 2019

Recall

Recall Number
Z-0278-2020
Event Number
83852
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
KDI
Status
Open, Classified
Root Cause
Software design
Initiated
October 4, 2019
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

Prismaflex Control Unit, software versions below 7.21

Reason

Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.

Action

Baxter Healthcare notified customers on about 10/04/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that Prismaflex Control Units with software versions 7.20 and below have the potential to produce communication errors. Instructions included that operators may continue to use the control units until the upgrade is performed, but if a communication error alarm does occur, follow the instructions presented on the graphical user interface and/or in the operators manual. It must be reinforced that it is important to manually return the extracorporeal blood to the patient, and re-train users on this process, outlined on page 10:57 of the operators manual, if necessary. A local Baxter service representative will contact your facility to determine the correction plan and schedule the software upgrade. If the affected devices was purchased from Baxter, complete the provided Baxter Customer Reply Form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to [email protected]. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. For a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures. Questions regarding this recall can be directed to Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.

Distribution

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. International distribution to Australia, Bangladesh, Brunei, China, Hongkong, Indonesia, India, Japan, New Zealand, Taiwan, Korea, Singapore, Malaysia, Macau, Thailand, Vietnam, Canada, Argentina, Costa Rica, Cuba, Dominican Republic, Haiti, Barbados, Venezuela, St. Kitts, and EMEA.

Quantity

988 units