FDA Recall Terminated

Automated External Defibrillator - HeartStart MRx Monitor/Defibrillator

Recall: Z-0278-2018 · Initiated September 28, 2017

Recall

Recall Number
Z-0278-2018
Event Number
78698
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
MSX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 28, 2017
Terminated
October 12, 2018
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Automated External Defibrillator - HeartStart MRx Monitor/Defibrillator

Reason

Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).

Action

The Philips Urgent Medical Device Correction notification/ Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of providing a hardware correction, free of charge, for all units affected by this issue.

Distribution

20 units affected. 18 domestic, 2 foreign.

Quantity

20 units ( 18 domestic, 2 foreign)