FDA Recall Terminated

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Recall: Z-0278-2013 · Initiated October 16, 2012

Recall

Recall Number
Z-0278-2013
Event Number
63344
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DWC
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 16, 2012
Posted
November 13, 2012
Terminated
July 11, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Reason

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Action

TERUMO sent an Urgent Medical Device Recall letter dated October 15, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Terumo has determined that a field correction is required to notify affected customers that the boards may fail, instruct the users that in a case of failure to swich to the second channel on the device or use a back-up unit, and that TCVS will replace the affected boards. Customers with questions were instructed to call 1-800-621-2818. For questions regarding this recall call 734-741-6173.

Distribution

Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI

Quantity

19 units