FDA Recall Terminated

OTC Pregnancy Tests. Name: Sure Check hCG Cassette Pregnancy Test. --- Description: the Sure Check hCG Cassette Pregnancy test consists of a lateral flow membrane pad containing anti- hCG monoclonal antibody conjugated to colloidal gold contained in a plastic cassette. It differs from the Sure Check Midstream product in that the lateral flow membrane pad is placed in a cassette and the urine sample is added by pipette rather than by urinating on the absorbent wick. The device is packaged in a foil pouch with desiccant. This product is shipped as bulk packaged product of varying quantities (Catalog # PT127-OTC ordered by Innovative Chemical Co., and Pruebas Moderna Del Istmo; Catalog # PT210 ordered by LABSCO). --- 510(k) K933529/A.

Recall: Z-0277-05 · Initiated October 6, 2004

Recall

Recall Number
Z-0277-05
Event Number
30288
Firm
Chembio Diagnostic System Inc
FEI Number
2431980
Product Code
LCX
Status
Terminated
Root Cause
Other
Initiated
October 6, 2004
Posted
December 3, 2004
Terminated
March 24, 2005
Address
3661 Horseblock Road, Medford, NY, 11763-2215

Description

OTC Pregnancy Tests. Name: Sure Check hCG Cassette Pregnancy Test. --- Description: the Sure Check hCG Cassette Pregnancy test consists of a lateral flow membrane pad containing anti- hCG monoclonal antibody conjugated to colloidal gold contained in a plastic cassette. It differs from the Sure Check Midstream product in that the lateral flow membrane pad is placed in a cassette and the urine sample is added by pipette rather than by urinating on the absorbent wick. The device is packaged in a foil pouch with desiccant. This product is shipped as bulk packaged product of varying quantities (Catalog # PT127-OTC ordered by Innovative Chemical Co., and Pruebas Moderna Del Istmo; Catalog # PT210 ordered by LABSCO). --- 510(k) K933529/A.

Reason

Poor sealing of pouches and degradation of products. Firm received an e-mail from a distributor of an increase in complaints. In response to the complaints, the distributor conducted its own functional testing and noticed a degradation in performance of the two lots. Chembio's investigation revealed invalid results (no control lines) and some false negatives.

Action

Recall letters on 10/6/04, followed by a second letter on/about 10/20/04. Recall letters mailed on 11/9/04 to cover recall expansion for 4 additional lots.

Distribution

Distributors in KS, TN, NC, MA, NY, KY, FL, NV, and Puerto Rico for further distribution to their retail customers. Products also distributed to accounts in Barbados WI, Israel, Costa Rica and Canada.

Quantity

42,005 (total of 2 lots)