FDA Recall
Terminated
Spacelabs flat panel monitor, Model 91415-A
Recall: Z-0269-04
·
Initiated October 29, 2003
Recall
- Recall Number
- Z-0269-04
- Event Number
- 27598
- Firm
- Datex-Ohmeda, Inc.
- FEI Number
- 3010157426
- Product Code
- DXJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 29, 2003
- Posted
- July 20, 2004
- Terminated
- September 17, 2004
- Address
- 5150 220th AVE SE, Issaquah, WA, 98029
Description
Spacelabs flat panel monitor, Model 91415-A
Reason
Potential for power supply to fail resulting in failure of the audio (alarm tones) to fail.
Action
A letter dated October 31, 2003 was sent on October 29-31, 2003 to all US customers via FED-EX. An ORDER FORM accompanied the letter for consignees to request a replacement power supply or to indicate they did not have the affected power supply.
Distribution
The firm distributes to hospitals nationwide. The firm also distributes internationally.
Quantity
2734 units for domestic consignees and 1099 for international consignees