VITROS chemistry Products Cl- Slides, Catalog No. 8445207 (which does NOT support Urine CL-). MicroSlides which contain reagents in a dry, multi-layered form.

Recall: Z-0268-2020 · Initiated September 10, 2019

Recall

Recall Number
Z-0268-2020
Event Number
83879
Firm
Ortho-Clinical Diagnostics
FEI Number
1000305840
Product Code
CGZ
Status
Terminated
Root Cause
Process control
Initiated
September 10, 2019
Terminated
November 13, 2020
Address
513 Technology Blvd, Rochester, NY, 14626-3601

Description

VITROS chemistry Products Cl- Slides, Catalog No. 8445207 (which does NOT support Urine CL-). MicroSlides which contain reagents in a dry, multi-layered form.

Reason

Users may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge.

Action

On September 10, 2019, the firm distributed Urgent Product Correction Notification letters to affected customers. Customers were asked to take the following actions: "Complete and return the Confirmation of Receipt form. "If you experience a slide dispense issue, try to reposition the ABP by following the instructions in the enclosed Resolving Slide Dispense Issues document. Please place the instructions by each VITROS System in your facility that processes Microslides. "If you are unable to reposition the ABP, remove the affected cartridge and replace it with a new one. Complete and return the Request for Credit form as needed to receive credit for the slides remaining within affected and discarded cartridges. "Please forward this notification if the product was distributed outside of your facility. The firm will provide credit for the slides remaining within affected cartridge for which you are unable to reposition the ABP. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.

Distribution

Domestic distribution nationwide. Foreign distribution to Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, the Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Quantity

1365545 total