FDA Recall Open, Classified

Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840

Recall: Z-0267-2022 · Initiated October 16, 2021

Recall

Recall Number
Z-0267-2022
Event Number
88877
Firm
Western/Scott Fetzer Company
FEI Number
1526809
Product Code
ECX
Status
Open, Classified
Root Cause
Device Design
Initiated
October 16, 2021
Address
875 Bassett Rd, Westlake, OH, 44145-1142

Description

Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840

Reason

VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke

Action

Western / Scott Fetzer Co. issued Urgent Medical device Recall letter, dated 10/15/21 distributed via USPS to consignees on 10/16/21, and emailed to the same consignees on 10/18/21. Letter states reason for recall , health risk and action to take: Fill facility employees are asked to verify the functionality of each unit through the observation of information on the digital display. Note that this direction and expectation is already clearly indicated in the product Operating Instructions and Inspection Guidelines. If the digital display shows the expected information as depicted in the Operating Instructions and Inspection Guidelines, the unit is functioning properly and may remain in use. If the digital display is blank, the unit is to be isolated and returned to Western for inspection. As a longer term solution, beginning in October 2021, product within the scope of the recall is to be corrected as follows: 1. For Digital VIPR products that do not include a cylinder (MNDR-600), units will be returned to Western via RGA to perform the following services: " Remove shroud from unit to determine battery type used " If 3500mAh battery found, replace it with 4000mAh battery " If 4000mAh battery found, inspect for battery lead wire damage, replace if damage found " Remove potential burr / sharp edge from screw boss on regulator body " Replace back shroud with modified back shroud (pinch points removed) " Reassemble unit per revised assembly procedure " Apply new control knob sticker (green logo, silver background) Page 3 of 6 2. For Digital products that include a cylinder (ROC-9840 and MNDS-602), Western employees will visit customer fill facility locations to perform the following services: " Remove shroud from unit to determine battery type used " If 3500mAh battery found, replace it with 4000mAh battery " If 4000mAh battery found, inspect for battery lead wire damage, replace if damage found " Remove potential burr / sharp edge from screw boss on regulator body " Re

Distribution

CT, MD, MT, OH

Quantity

11341 units