VITROS chemistry Products Cl- Slides, Catalog No. 6844471 (which supports Urine CL-). MicroSlides which contain reagents in a dry, multi-layered form.
Recall
- Recall Number
- Z-0267-2020
- Event Number
- 83879
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000305840
- Product Code
- CGZ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 10, 2019
- Terminated
- November 13, 2020
- Address
- 513 Technology Blvd, Rochester, NY, 14626-3601
Description
VITROS chemistry Products Cl- Slides, Catalog No. 6844471 (which supports Urine CL-). MicroSlides which contain reagents in a dry, multi-layered form.
Users may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge.
On September 10, 2019, the firm distributed Urgent Product Correction Notification letters to affected customers. Customers were asked to take the following actions: "Complete and return the Confirmation of Receipt form. "If you experience a slide dispense issue, try to reposition the ABP by following the instructions in the enclosed Resolving Slide Dispense Issues document. Please place the instructions by each VITROS System in your facility that processes Microslides. "If you are unable to reposition the ABP, remove the affected cartridge and replace it with a new one. Complete and return the Request for Credit form as needed to receive credit for the slides remaining within affected and discarded cartridges. "Please forward this notification if the product was distributed outside of your facility. The firm will provide credit for the slides remaining within affected cartridge for which you are unable to reposition the ABP. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Domestic distribution nationwide. Foreign distribution to Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, the Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
1365545 total