FDA Recall Terminated

Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack. The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes, as well as clarity.

Recall: Z-0266-2022 · Initiated October 18, 2021

Recall

Recall Number
Z-0266-2022
Event Number
88879
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
KQO
Status
Terminated
Root Cause
Software design
Initiated
October 18, 2021
Terminated
May 31, 2023
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack. The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes, as well as clarity.

Reason

A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinine levels up to 9,000 mg/L.

Action

On October 18, 2021, the firm, Roche, sent "URGENT MEDICAL DEVICE CORRECTION" letters to affected consignees. Customers were notified of the potential for false negative nitrite test results with elevated endogenous creatinine concentrations. Customers were provided with an amended cobas u pack method sheet that provides the updated interference claim for endogenous creatinine. Nitrite testing is commonly included in the process of diagnosing urinary tract infections. In case of false negative nitrite results, particularly in asymptomatic patients or patient with nonspecific symptoms, further diagnostic measures in urine might be skipped, resulting in an increased potential risk for disease progression and complications of UTI. Actions required: - Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. - Review the updated cobas u pack method sheet included with the letter and update your laboratory's procedures, if needed. - Complete and return the customer response form via email to [email protected]. - File the recall notice for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.

Distribution

US Nationwide distribution.

Quantity

636