FDA Recall Terminated

Rapid Strand Rx; the RAPID Strand Rx kit consists of custom loaded absorbable seeding spacers and I-125 radionuclide brachytherapy sources (seeds) spaced at prescribed distance and configuration within a sleeve (tube) made of absorbable suture material, stiffened, loaded into prostate seeding needles, packaged and then sterilized by Gamma sterilization method, accompanied by a calibration strand, a sealed source calibration certificate (100% assay report), a radiograph of the loaded needles in the needle-holding card, and a source certification technical data sheet.; Products Provided by: Oncura, 3350 North Ridge Avenue, Arlington Heights, IL 60004 USA, Prescription Filled by: AnazaoHealth, 5710 Hoover Blvd., Tampa, FL 33634 USA; product identifier: I125RSRX, model 7000. RAPID Strand Rx is indicated for permanent interstitial implantation of selected localized tumors that are of low to moderate radio sensitivity. It may be used either as primary treatment (such as prostate cancer or unresectable tumors) or for the treatment of residual disease after excision of the primary tumor. RAPID Strand Rx may be indicated for use concurrent with or at the completion of other treatment modalities such as external beam radiation therapy.

Recall: Z-0266-2012 · Initiated September 20, 2011

Recall

Recall Number
Z-0266-2012
Event Number
60005
Firm
Medi-Physics Inc. dba GE Healthcare
FEI Number
1417338
Product Code
KXK
Status
Terminated
Root Cause
Process control
Initiated
September 20, 2011
Posted
November 22, 2011
Terminated
January 20, 2012
Address
3350 N Ridge Ave, Arlington Heights, IL, 60004-1412

Description

Rapid Strand Rx; the RAPID Strand Rx kit consists of custom loaded absorbable seeding spacers and I-125 radionuclide brachytherapy sources (seeds) spaced at prescribed distance and configuration within a sleeve (tube) made of absorbable suture material, stiffened, loaded into prostate seeding needles, packaged and then sterilized by Gamma sterilization method, accompanied by a calibration strand, a sealed source calibration certificate (100% assay report), a radiograph of the loaded needles in the needle-holding card, and a source certification technical data sheet.; Products Provided by: Oncura, 3350 North Ridge Avenue, Arlington Heights, IL 60004 USA, Prescription Filled by: AnazaoHealth, 5710 Hoover Blvd., Tampa, FL 33634 USA; product identifier: I125RSRX, model 7000. RAPID Strand Rx is indicated for permanent interstitial implantation of selected localized tumors that are of low to moderate radio sensitivity. It may be used either as primary treatment (such as prostate cancer or unresectable tumors) or for the treatment of residual disease after excision of the primary tumor. RAPID Strand Rx may be indicated for use concurrent with or at the completion of other treatment modalities such as external beam radiation therapy.

Reason

There is a potential brachytherapy over or under dose with the RAPID Strand Rx kit that may impact patient therapy.

Action

GE Healthcare/Oncura sent a "Urgent Medical Device Correction" letter dated September 20, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were advised to assess the pre and post plan dosimetry for their patient, taking into account the possible variation of the seed strength. GE Healthcare/Oncura will contact the customers in the near future to confirm their receipt of the letter and to discuss any questions they may have. Please call if you have any further questions (847) 398-8400 and 011-44-1494-737251 (Amersham, United Kingdom, GMT)

Distribution

Worldwide Distribution -- USA (nationwide) including the States of Arizona, Florida, Louisiana, Mississippi, New Jersey, New York, Pennsylvania, Puerto Rico, Virginia and Washington, and the countries of Canada and the United Kingdom

Quantity

23 orders