FDA Recall Terminated

ETEST OXACILLIN OX 256 WW F100 In vitro diagnostic, a quantitative technique for determining the antimicrobial susceptibility of bacteria species.

Recall: Z-0265-2018 · Initiated April 13, 2017

Recall

Recall Number
Z-0265-2018
Event Number
78688
Firm
BioMerieux SA Chemin De L'Orme Marcy L'Etoile France
FEI Number
1000611339
Product Code
JWY
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 13, 2017
Posted
December 19, 2017
Terminated
August 18, 2020

Description

ETEST OXACILLIN OX 256 WW F100 In vitro diagnostic, a quantitative technique for determining the antimicrobial susceptibility of bacteria species.

Reason

QC failures (MIC out of range high) for S.aureus ATCC 29213 strain on ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) reported. The product behavior could lead to False Resistant results.

Action

BioMerieux sent an Urgent Field Safety Notice on April 30, 2017. Short-term Action: Subsidiaries and distributors are required to discard all ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) under (PSS#3320) remaining in their inventory. In order to avoid any backorders on the field during the transition to the SPB configuration that is conformed to the expected specifications, Subsidiaries and Distributors were instructed to distribute the Urgent Field Safety Notice (customer letter) to inform Medical customers and Veterinary customers that have received and will receive the product. The Urgent Field Safety Notice informed customers of the non-conformity along with instructions to continue the use of the impacted products with recommendations described in customer letter, and in addition, for the Medical Customers, to discuss any concerns they may have regarding previously reported Susceptible results on the impacted batches Listed Table 1 with their Laboratory Medical Director to determine the appropriate course of action. Long-term Action: A CAPA (PR#1101306) has been initiated in order to identify the root-cause and prevent the issue to recur. They were instructed to complete and return the attached Acknowledgement Form by Fax to confirm receipt of this notice. For further questions, please call (314) 731-8694.

Distribution

Worldwide Distribution - USA (nationwide) and to the countries of : Algeria, Austria, Australia, Belarus, Bosnia-Herz., Canada, Croatia, Czech Republic, Egypt, Estonia, France, Georgia, Germany, Greece, Guam, Hong Kong, Hungary, India, Israel, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Nigeria, Peru, Philippines, Qatar, Russian Fed., Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Utd.Arab Emir.

Quantity

1585 units