FDA Recall Terminated

Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer.

Recall: Z-0265-05 · Initiated November 8, 2004

Recall

Recall Number
Z-0265-05
Event Number
30421
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
CGN
Status
Terminated
Root Cause
Other
Initiated
November 8, 2004
Posted
November 30, 2004
Terminated
August 13, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer.

Reason

A rare condition of the Access 2 Immunoassay Systems software versions 2.0, 2.1 and 2.2 could present a risk of an erroneous confirmatory result under a specific set of circumstances. The Chlamydia Blocking results may be affected. The instrument will erroneously perform the blocking calculation on the affected test by using the mean of the diluted and neat results.

Action

Product Corrective Action (PCA) letter that informs customers (who ordered Access Chlamydia reagent in the past 24 months and have an Access 2 Immunoassay System) of this potential issue sent via US mail on November 8, 2004. The letter instructs customers to use unique Sample IDs for all samples assayed on the system - including those which have been manipulated offline prior to assay. A Technical Information Update will be issued to inform all support personnel of this issue.

Distribution

Nationwide and Canada

Quantity

1,197