Permobil TiLite, Models: Aero Z, and ZRA

Recall: Z-0264-2022 · Initiated September 24, 2021

Recall

Recall Number
Z-0264-2022
Event Number
88718
Firm
TiSport, Llc
FEI Number
3002653226
Product Code
IOR
Status
Open, Classified
Root Cause
Component change control
Initiated
September 24, 2021
Address
2701 W Court St, Pasco, WA, 99301-3911

Description

Permobil TiLite, Models: Aero Z, and ZRA

Reason

Some affected wheelchairs, with transit tie downs, that are used as a passenger seat in a motor vehicle may not be configured in the right orientation to perform as intended if involved in a motor vehicle accident. Some affected wheelchairs can no longer be used as a passenger seat until transit tie downs are replaced with redesigned transit tie downs, verified, and released to the market.

Action

On 09/24/21, distribution of correction notices started, via mail, to distributors. Distributors were advised the following: For AERO Z wheelchairs, the transit tie downs should have an inward orientation with transit bracket in rear most position. The wheelchair provider was provided with clear and thorough work instructions on how to inspect and, when necessary, correctly configure the transit tie downs. Stop using TiLite ZRA with transit tie downs as a passenger seat in a motor vehicle. Remove transit tie downs on TiLite ZRA wheelchairs. The recalling firm, if requested, will supply at no cost to the provider or the end-user an Aero Z with transit tie downs as a replacement for the ZRA. Once transit tie downs for the ZRA have been redesigned, verified, and released to the market, the updated transit tie downs will be provided, and the return of the replacement Aero Z will be coordinated. On October 1st, 2021, an update will be provided on our progress regarding the redesigned transit tie downs for ZRA wheelchair models. Customers with questions or concerns are encouraged to email [email protected] or phone 800 736 0925, option 2. Distributors were advised that this market correction needs to be carried out at the consumer and end-user level. They were asked to immediately identify and contact their customer(s) with affected serial numbers. Distributor notification to customers may be more effective by including a copy of the end-user letter, which is attached with the distributor correction notice. If they are unable to contact the end-user or the transit tie downs/wheelchair are no longer available, please indicate this on the Serial Number Verification Form.

Distribution

US Nationwide Distribution: MD, PA, VA, IN, NY, MO, AR, NE, NJ, WA, NM, IL, CA, OH, KY, LA, OK, WI, TX, NC, SD, MN, MA, OR; OUS (Foreign): AUSTRALIA, CANADA, FRANCE, GERMANY, IRELAND, ITALY, UNITED KINGDOM, URUGUAY, NEW ZEALAND, IRELAND, SWEDEN

Quantity

5373