Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Recall: Z-0263-2016 · Initiated September 14, 2015

Recall

Recall Number
Z-0263-2016
Event Number
72171
Firm
NeuroLogica Corporation
FEI Number
3004938766
Product Code
IZI
Status
Terminated
Root Cause
Device Design
Initiated
September 14, 2015
Posted
November 10, 2015
Terminated
January 19, 2017
Address
14 Electronics Ave, Danvers, MA, 01923-1011

Description

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Reason

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Action

An Urgent - Field Safety Notice identifying the affected products, problem involved, root cause, and potential hazards, was sent to customers. In the event the issue occurs, users can release the drive handle or press the emergency switch and the unit will stop. Then, reboot the system. If unintended movements occur, please contact local representative. To correct the issue, service engineers will visit to apply the corrective action. For further information or support concerning this problem, contact local representative.

Distribution

Distributed to the states of NC, OH, MA, and MO.

Quantity

7 devices