SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465

Recall: Z-0262-5 · Initiated October 25, 2004

Recall

Recall Number
Z-0262-5
Event Number
30362
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
CHI
Status
Terminated
Root Cause
Other
Initiated
October 25, 2004
Posted
November 30, 2004
Terminated
June 17, 2005
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465

Reason

Beckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.

Action

The week of 10/25/2004, a product corrective action (PCA) letter was mailed vis U.S. mail to all customers shipped LIPA reagent for the 24 months preceding the discovery of the issue. Customers instructed to repeat any sample that produces a lipase value of > 170 U/L that cannot be confirmed by patient history.

Distribution

US and Canada

Quantity

6,698