FDA Recall
Terminated
SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465
Recall: Z-0262-5
·
Initiated October 25, 2004
Recall
- Recall Number
- Z-0262-5
- Event Number
- 30362
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- CHI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 25, 2004
- Posted
- November 30, 2004
- Terminated
- June 17, 2005
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465
Reason
Beckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.
Action
The week of 10/25/2004, a product corrective action (PCA) letter was mailed vis U.S. mail to all customers shipped LIPA reagent for the 24 months preceding the discovery of the issue. Customers instructed to repeat any sample that produces a lipase value of > 170 U/L that cannot be confirmed by patient history.
Distribution
US and Canada
Quantity
6,698