FDA Recall Terminated

Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.

Recall: Z-0262-2018 · Initiated July 28, 2017

Recall

Recall Number
Z-0262-2018
Event Number
77797
Firm
NXTHERA
FEI Number
3007795803
Product Code
KNS
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 28, 2017
Terminated
August 1, 2018
Address
7351 Kirkwood Ln N, Maple Grove, MN, 55369-5201

Description

Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.

Reason

Needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.

Action

Customers were notified initially by sales representatives on or about July 18, 2017. Any affected devices were to be returned to NxThera. Additionally, a formal recall letter was mailed to customers on approximately July 28, 2017. NxThera's telephone contact information and FEDEX number will be provided on the verification form. For further questions, please call (763) 515-2084..

Distribution

USA Distribution to the states of : AL, AZ, CA, CO, FL, GA, IL, MA, MI, MN, MO, NE, NJ, NY, OH, PA, TN, UT, WA.

Quantity

113 units total