GE Healthcare, SIGNA HDxt 3.0T. Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner for use as a diagnostic imaging device.
Recall
- Recall Number
- Z-0259-2016
- Event Number
- 72301
- Firm
- GE Healthcare
- FEI Number
- 2126677
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 31, 2015
- Posted
- November 10, 2015
- Terminated
- September 7, 2018
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Healthcare, SIGNA HDxt 3.0T. Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner for use as a diagnostic imaging device.
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
GE Healthcare sent an Urgent Medical Device Correction letter dated August 31, 2015. The letter was addressed to Hospital Administrators/Risk Managers, Radiology Department Managers, & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.
215