DJO EMPOWR Knee Punch Handle, REF: 801-05-040
Recall
- Recall Number
- Z-0258-2024
- Event Number
- 93214
- FEI Number
- 1000116912
- Product Code
- MBH
- Status
- Open, Classified
- Root Cause
- Process change control
- Initiated
- September 29, 2023
- Posted
- November 7, 2023
- Address
- 9800 Metric Blvd, Encore Medical, LP, Austin, TX, 78758-5445
Description
DJO EMPOWR Knee Punch Handle, REF: 801-05-040
Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.
On 9/29/23, recall notices were emailed to customers who were asked to do the following: 1) Knee punch handles should be visually inspected prior to use for any damage. Upon completion of surgery, visually inspect and verify there is no breakage. The attached quality bulletin provides guidance on an acceptable and unacceptable device. 2) Contact your customer service rep at [email protected] to report failures or to request replacements. 3) Complete and return the response form that asks customers to check all related devices in inventory and report the number of related devices in inventory. Questions can be directed to [email protected]
US: VA, TN, MN, TX, IL, IN, KY, OH, SC, MS, RI, ME, NJ, NY, MD, CA, KS, LA, PR, AL, FL, GA, AZ, WA