FDA Recall Terminated

Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure Interface Cable (Part # 3006314) for use with Abbott/Hospira IBP Transducer b) Invasive Blood Pressure Interface Cable (Part # 3008772) for use with Braun IBP Transducers. The product is an accessory to patient physiological monitor.

Recall: Z-0258-2007 · Initiated September 29, 2006

Recall

Recall Number
Z-0258-2007
Event Number
36596
Firm
Medrad Inc
FEI Number
2520313
Product Code
MWI
Status
Terminated
Root Cause
Other
Initiated
September 29, 2006
Posted
December 15, 2006
Terminated
June 20, 2007
Address
1 Medrad Dr, Indianola, PA, 15051-9759

Description

Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure Interface Cable (Part # 3006314) for use with Abbott/Hospira IBP Transducer b) Invasive Blood Pressure Interface Cable (Part # 3008772) for use with Braun IBP Transducers. The product is an accessory to patient physiological monitor.

Reason

Interference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes the invasive blood pressure readings on the physiological monitor to display inaccurate readings.

Action

The recalling firm contacted all affected customers via telephone or letter dated November 20, 2006 to inform them of the problem and instruct them to return the cables or apply corrective labeling.

Distribution

Worldwide distribution ---- including USA state of CA, CO, HI, IL, IN, MA, MD, NJ, NY, OH, PA, TX, and WA and countries of The Netherlands, Denmark, Australia, Brazil, Italy, and Japan.

Quantity

80 units