FDA Recall Terminated

Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

Recall: Z-0256-2013 · Initiated June 1, 2012

Recall

Recall Number
Z-0256-2013
Event Number
62580
Firm
Sysmex America, Inc.
FEI Number
1000515253
Product Code
GKZ
Status
Terminated
Root Cause
Device Design
Initiated
June 1, 2012
Posted
November 6, 2012
Terminated
August 22, 2016
Address
1 Nelson C White Pkwy, Mundelein, IL, 60060-9528

Description

Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

Reason

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Action

Sysmex America, Inc. sent a "PRODUCT NOTIFICATION" letter dated May 2012 to all affected customers on June 1, 2012. The letter identifies the product, problem, and actions to be taken. The letter instructs customers to review patient results to determine if their Reference Intervals need adjustment, and to review QC results via Insight and document any change that may be linked to the date of calibration if appropriate. Any questions concerning the information contained in the notification were directed to the Sysmex Technical Assistance Center at 1-888-879-7639.

Distribution

Nationwide distribution: USA including Puerto Rico.

Quantity

69 units