NEMIO Ultrasound System, Model SSA-550A , Software versions: V3.1.000B, V3.5.0, V3.5.000A, V3.5.000B, V3.6.0, V3.6.000A, V4.0.0, V4.0.000A, V4.l.0, V4.l.000A, V4.2.0, V4.3.0, V4.5.0, V4.5.000A, Toshiba America Medical Systems, Inc., Tustin, CA.
Recall
- Recall Number
- Z-0253-2008
- Event Number
- 44811
- Firm
- Toshiba American Med Sys Inc
- FEI Number
- 2020563
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 29, 2007
- Posted
- November 17, 2007
- Terminated
- April 5, 2012
- Address
- 2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068
Description
NEMIO Ultrasound System, Model SSA-550A , Software versions: V3.1.000B, V3.5.0, V3.5.000A, V3.5.000B, V3.6.0, V3.6.000A, V4.0.0, V4.0.000A, V4.l.0, V4.l.000A, V4.2.0, V4.3.0, V4.5.0, V4.5.000A, Toshiba America Medical Systems, Inc., Tustin, CA.
Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than the true TI value, and (3) Manuals contain incorrect values for MI/TI values.
Urgent Medical Device Notification letters were sent to customers on August 29, 2007. Customers were notified of the several issues related to the Nemio SSA-550A and Nemio SSA-580A recall and asked to share this information with all users of the device. Customers were informed that Toshiba is preparing software and corrected manuals for distribution to address these issues. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department.
Nationwide
1,278 units