FDA Recall Terminated

NEMIO Ultrasound System, Model SSA-550A , Software versions: V3.1.000B, V3.5.0, V3.5.000A, V3.5.000B, V3.6.0, V3.6.000A, V4.0.0, V4.0.000A, V4.l.0, V4.l.000A, V4.2.0, V4.3.0, V4.5.0, V4.5.000A, Toshiba America Medical Systems, Inc., Tustin, CA.

Recall: Z-0253-2008 · Initiated August 29, 2007

Recall

Recall Number
Z-0253-2008
Event Number
44811
Firm
Toshiba American Med Sys Inc
FEI Number
2020563
Product Code
IYO
Status
Terminated
Root Cause
Software design
Initiated
August 29, 2007
Posted
November 17, 2007
Terminated
April 5, 2012
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068

Description

NEMIO Ultrasound System, Model SSA-550A , Software versions: V3.1.000B, V3.5.0, V3.5.000A, V3.5.000B, V3.6.0, V3.6.000A, V4.0.0, V4.0.000A, V4.l.0, V4.l.000A, V4.2.0, V4.3.0, V4.5.0, V4.5.000A, Toshiba America Medical Systems, Inc., Tustin, CA.

Reason

Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than the true TI value, and (3) Manuals contain incorrect values for MI/TI values.

Action

Urgent Medical Device Notification letters were sent to customers on August 29, 2007. Customers were notified of the several issues related to the Nemio SSA-550A and Nemio SSA-580A recall and asked to share this information with all users of the device. Customers were informed that Toshiba is preparing software and corrected manuals for distribution to address these issues. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department.

Distribution

Nationwide

Quantity

1,278 units