STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Recall: Z-0252-2013 · Initiated June 15, 2012

Recall

Recall Number
Z-0252-2013
Event Number
63270
Firm
Diagnostica Stago, Inc.
FEI Number
2245451
Product Code
JPA
Status
Terminated
Root Cause
Software design
Initiated
June 15, 2012
Posted
November 6, 2012
Terminated
July 19, 2013
Address
5 Century Dr, Parsippany, NJ, 07054-4607

Description

STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Reason

Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite.

Action

Diagnostica Stago sent an Urgent Field Safety Notice dated June 12, 2012, to all affected customers via email return receipt requested on June 15, 2012. Customers were instructed to follow the instructions in the notification and complete the Customer Verification form and return to the Hotline by fax to 973-644-9348. For additional information, customers were instructed to contact the Hotline at 800-725-0607. For questions regarding this recall call 973-631-1200, ext 2044.

Distribution

Worldwide Distribution - USA (nationwide) and Canada

Quantity

232 units US; 5 units Canada