FDA Recall Terminated

Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<25K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.

Recall: Z-0250-2014 · Initiated September 16, 2013

Recall

Recall Number
Z-0250-2014
Event Number
66306
Firm
Carestream Health Inc.
FEI Number
1315356
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
September 16, 2013
Posted
November 13, 2013
Terminated
December 21, 2015
Address
150 Verona St, Rochester, NY, 14608-1733

Description

Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<25K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.

Reason

Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.

Action

Carestream sent an Urgent Medical Device Recall letter dated September 13, 2013, to all affected customers via Fed Ex on September 16, 2013, informing them that a software problem may have the potential to occur under certain conditions when using the CARESTREAM Client Workstation. When certain conditions are met, it appears that the child background process completes its tasks before the parent process and as a result the parent process considers its work done and terminates. In the event that a software problem does occur, the scan may not be fluent or the image numbers may not be sequential. A software patch is currently being deployed by Carestream Health. Customers were instructed that they would be contacted by a Carestream Service Engineer to coordinate the deployment. Customers with questions were instructed to call 1-866 737 2487 or they may call the Carestream Customer Care Center at 1-800-328-2910, 8AM through 8pm Eastern Standard Time. For questions regarding this recall call 1-800-328-2910.

Distribution

Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, IA, IL, MA, MD, ME, MN, MO, NC, NH, NV, NY, OH, OK, PA, RI, TN, UT, VA WA, and WI and Internationally to Canada, Japan, and Kuwait.