Solar 9500 Information Monitor with software version 3A.

Recall: Z-0250-04 · Initiated November 10, 2003

Recall

Recall Number
Z-0250-04
Event Number
27811
Firm
General Electric Medical Systems Information Technology
FEI Number
2124823
Product Code
MHX
Status
Terminated
Root Cause
Other
Initiated
November 10, 2003
Posted
December 18, 2003
Terminated
November 23, 2005
Address
8200 W Tower Ave, Milwaukee, WI, 53223-3219

Description

Solar 9500 Information Monitor with software version 3A.

Reason

Software gives inaccurate readings when transferring TRAM module from Solar 9500 to other Solar monitoring systems without discharging TRAM first.

Action

A letter dated November 10, 2003 was sent to the consignees. The letter recommended certain actions to avoid the problems and provided options for the software to be updated.

Distribution

The product was distributed to consignees located nationwide in the United States and worldwide.

Quantity

maximum 648 Solar 9500s with Version 3A software (297 Solar 9500 Vers. 3A software were manufactured plus 131 software update kits to update multiple)