FDA Recall Open, Classified

3M Steri-Drape, Small Towel Drape, REF 1000NSD, general surgery drape

Recall: Z-0249-2023 · Initiated October 6, 2022

Recall

Recall Number
Z-0249-2023
Event Number
90989
Firm
3M Healthcare Business
FEI Number
2126770
Product Code
PUI
Status
Open, Classified
Root Cause
Component change control
Initiated
October 6, 2022
Posted
November 18, 2022
Address
2510 Conway Ave E, Saint Paul, MN, 55144-4147

Description

3M Steri-Drape, Small Towel Drape, REF 1000NSD, general surgery drape

Reason

During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.

Action

The firm began disseminating an URGENT Product Recall notice their consignees by email or letter beginning 10/06/2022. The notice explained the issue and requested the following: Distributors: "Actions Requested: 1. Immediately discontinue distribution of affected models and lots of 3M" Steri-Drape" Surgical Drapes. Reference Attachment 1 for list of models and lots. 2. Determine if any stock of the affected models and lots remain in inventory. 3. Complete the Distributor Acknowledgement Form with the following: " Acknowledge that you have read and understood this letter whether or not you have any 3M" Steri-Drape" Surgical Drapes in stock at your facility; and " Indicate if you have any product in stock and dispose of this material per facility policy. " Email the completed form to [email protected]. 4. Notify any of your customers who have purchased this product using the Customer Letter and Customer Acknowledgement Form provided. 5. To request credit, contact 3M Health Care Customer Helpline at 1-800-228-3957, option 7 or by email to [email protected]." Direct accounts: "Actions Requested: 1. Immediately cease use of affected models and lots of 3M" Steri-Drape" Surgical Drapes. Reference Attachment 1 for list of models and lots. 2. Determine if any stock of the affected models and lots remain in inventory. NOTE: if you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 3. Complete the Customer Acknowledgement Form with the following: " Acknowledge that you have read and understood this letter whether or not you have any 3M" Steri-Drape" Surgical Drapes in stock at your facility; and " Indicate if you have any product in stock and dispose of this material per facility policy. " Email the completed form to [email protected]. 4. To request credit, contact 3M Health Care Customer Helpline at 1-800-228-3957, option 7 or by email t

Distribution

Worldwide distribution.

Quantity

248,750 units