FDA Recall Terminated

Sling Rope/Connector Assembly Kits for the Sara and Sara Nova standing and raising aids; the kit is comprised of 2 x 200 cm ropes, 2 knobs, 2 sling connectors, 2 socket cap screws, 2 nylon nuts and 2 spacers; Arjo Inc., 50 N. Gary Avenue, Roselle, IL 60172; Model KS1006

Recall: Z-0248-2007 · Initiated October 30, 2006

Recall

Recall Number
Z-0248-2007
Event Number
36680
Firm
Arjo, Inc.
FEI Number
1419652
Product Code
FSA
Status
Terminated
Root Cause
Other
Initiated
October 30, 2006
Posted
December 6, 2006
Terminated
January 18, 2008
Address
50 Gary Ave, Ste A, Roselle, IL, 60172-1684

Description

Sling Rope/Connector Assembly Kits for the Sara and Sara Nova standing and raising aids; the kit is comprised of 2 x 200 cm ropes, 2 knobs, 2 sling connectors, 2 socket cap screws, 2 nylon nuts and 2 spacers; Arjo Inc., 50 N. Gary Avenue, Roselle, IL 60172; Model KS1006

Reason

The sling ropes were manufactured with insufficient rope length past the knot to prevent the knot from coming loose.

Action

Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that affected rope kits were incorrectly assembled, and could lead to the rope separating from the cone-shaped knob and causing the patient to fall back from the lift. The letters listed the specific quantities, order numbers and dates for the kits shipped to the customer, and requested the removal from use and return of the affected rope kits. Replacement rope kits were sent with the recall letters. Any questions were directed to Arjo''s Quality Department at 1-800-323-1245, ext. 6140.

Distribution

Nationwide, including Florida, Illinois, Indiana, Iowa, Kansas, Massachusetts, New York, Tennessee, Virginia and Wyoming.

Quantity

15 kits