TherMax Blood Warmer Unit
Recall
- Recall Number
- Z-0247-2020
- Event Number
- 83816
- Firm
- Baxter Healthcare Corporation
- FEI Number
- 1417572
- Product Code
- KDI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 19, 2019
- Posted
- October 30, 2019
- Terminated
- December 11, 2020
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
TherMax Blood Warmer Unit
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
Baxter Healthcare notified customers on about 09/19/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that specific TherMax Blood Warmers listed below may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. The affected units may not have an adequate ground path, which could cause the units to become damaged if used with improperly grounded outlets. The damage could cause the TherMax Blood Warmer units to be nonfunctional during clinical use. Instructions included that customers could continue to safely use affected devices with a properly grounded outlet, that a local Baxter representative will contact each facility to arrange for the correction of affected devices, to complete and return the Baxter Customer Reply Form via fax to 224-270-5457 or scanning and e-mailing it to [email protected], and to notify customers if the product was further distributed. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand
504 units total