FDA Recall Open, Classified

Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000

Recall: Z-0245-2023 · Initiated October 24, 2022

Recall

Recall Number
Z-0245-2023
Event Number
91000
Firm
Tollos
FEI Number
3008406502
Product Code
FNG
Status
Open, Classified
Root Cause
Equipment maintenance
Initiated
October 24, 2022
Posted
November 17, 2022
Address
8 Easter Ct, Ste J, Owings Mills, MD, 21117-3294

Description

Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000

Reason

Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).

Action

On October 24, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. The firm is updating their operator's manual and service manual with instructions to inspect the IC straps prior to use, along with photo illustrations of lifting straps demonstrating wear/damage conditions that indicate replacement must be performed. Access complete Operator s & Service Manuals on the web at www.tollos.com/service-bulletins. Tollos can be contacted by e-mail at [email protected].

Distribution

Worldwide - US Nationwide distribution and the countries of Canada and United Arab Emirates.

Quantity

1273 US; 70 OUS