FDA Recall Open, Classified

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

Recall: Z-0241-2026 · Initiated September 29, 2025

Recall

Recall Number
Z-0241-2026
Event Number
97682
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
OWB
Status
Open, Classified
Root Cause
Process control
Initiated
September 29, 2025
Posted
October 23, 2025

Description

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

Reason

Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.

Action

On September 29, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this notice to all users and follow the instructions in the letter. 2. Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). 3. Ensure the letter is in a place likely to be seen/viewed. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Starting in October 2025, Philips will replace the 11W control board of the affected Zenition 50 systems. Your local Philips representative will contact you to schedule a visit to perform this activity in your system.

Distribution

Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.

Quantity

29 units