Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.
Recall: Z-0241-06
·
Initiated November 22, 2004
Recall
- Recall Number
- Z-0241-06
- Event Number
- 31985
- Firm
- Stryker Howmedica Osteonics Corp.
- FEI Number
- 3003070421
- Product Code
- NDH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 22, 2004
- Posted
- December 1, 2005
- Terminated
- October 3, 2006
- Address
- 325 Corporate Dr, Mahwah, NJ, 07430-2002
Description
Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.
Reason
The Gamma Nail may have been manufactured with an incorrect lag screw hole angulation. Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees. Catalog number 3220-0320S, lot code K949408 was manufactured with a lag screw hole angulation of 125 degrees instead of 120 degrees.
Action
Notification letters and Product Accountability forms were sent via FedEx on 11/22/2004 with return receipt.
Distribution
The product was sent to Stryker Orthopaedics branches and agents throughout the US. There was one hospital that was a direct account.
Quantity
19