FDA Recall Terminated

Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.

Recall: Z-0241-06 · Initiated November 22, 2004

Recall

Recall Number
Z-0241-06
Event Number
31985
Firm
Stryker Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
NDH
Status
Terminated
Root Cause
Other
Initiated
November 22, 2004
Posted
December 1, 2005
Terminated
October 3, 2006
Address
325 Corporate Dr, Mahwah, NJ, 07430-2002

Description

Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.

Reason

The Gamma Nail may have been manufactured with an incorrect lag screw hole angulation. Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees. Catalog number 3220-0320S, lot code K949408 was manufactured with a lag screw hole angulation of 125 degrees instead of 120 degrees.

Action

Notification letters and Product Accountability forms were sent via FedEx on 11/22/2004 with return receipt.

Distribution

The product was sent to Stryker Orthopaedics branches and agents throughout the US. There was one hospital that was a direct account.

Quantity

19